Constrained optimized water suppression for 1H MRS of Metabolites and Macromolecules in 3 brain regions at 3T using sLASER
Contact: Kay Igwe, M.S. (kci2104 [at] columbia [dot] edu)
These data were collected as part of a study to test the efficacy of COWS, a new water suppression software package, for acquisition of both metabolite and macromolecule spectra. COWS sLASER localization.
Data types included are defaced T1-weighted 3D structural MRI images and single-voxel sLASER MRS data in the prefrontal cortex, posterior frontal lobe, and occipital lobe.
Ten healthy adults (5 women, 5 men, mean age = 31 ± 9 years).
Data from all 10 participants are available in this repository.
One 1-hr scan sessions for each participant with 2 T1 scans, one pre-MRS T1 scans used for voxel localization and a post-MRS T1 scan for measuring degreee of participant movement.
MR data were acquired at Zuckerman Mind Brain Behavior Institute at Columbia University on a MAGNETOM Prisma 3 T (Siemens Healthineers, Erlangen, Germany) using a 1H 32-reeive channel RF phased-array head and neck coil.
3D T1-weighted MPRAGE structural MRI acquisition parameters:
- MPRAGE
- TE/TR/TI = 2.26/2300/900 ms
- Flip angle = 8°
- Voxel resolution = 1 × 1 × 1 mm3
- FOV = 256 × 192 x 256 mm3
- Matrix size = 256 × 192 x 256
- Slices = 192
General acquisition parameters:
- Volumes of interest = prefrontal cortex, posterior frontal lobe, and occipital lobe
- Voxel resolution = 25 × 25 × 25 mm3
- Spectral width = 4000 Hz
- Number of points = 2048
- TE/TR = 26/2000 ms
- Number of transients (for Metabolite Scans) = 32
- Number of transients (for Macromolecule Scans) = 64
- Manual first-order B0 shimming using the Siemens Prisma VE11C native active shimming tool
MRS:
- Metabolite acquisition: Ace, Ala, Asc, Asp, Ch, Cr, GABA, Glc, Glu, Gln, Gly, GPC, GSH, Lac, mI, NAA, NAAG, PCr, PCh, PE, sI, and Tau
- Macromolecule ONLY acquisition
- Water reference acquisition
- Acquisition: sLASER
- COWS(7;236) water suppression: Water suppression scheme with 7 pulses that takes 236 ms to complete
- COWS(12;626) water suppression: Water suppression scheme with 12 pulses that takes 626 ms to complete
- VAPOR(12;626) water suppresion: Water suppression scheme with 7 pulses that takes 626 ms to complete
SPECTRA and MRI Per Participant
DAT FILE SPECTRA (RAW) Each *.dat file contains both the spectrum and associated water reference
- sub-*_acq*-01_svs_slaser_vapor7_metab_PFC.dat
- sub-*_acq*-02_svs_slaser_vapor7_mm_PFC.dat
- sub-*_acq*-03_svs_slaser_cows7_metab_PFC.dat
- sub-*_acq*-04_svs_slaser_cows7_mm_PFC.dat
- sub-*_acq*-05_svs_slaser_cows12_metab_PFC.dat
- sub-*_acq*-06_svs_slaser_vapor7_metab_Occipital.dat
- sub-*_acq*-07_svs_slaser_vapor7_mm_Occipital.dat
- sub-*_acq*-08_svs_slaser_cows7_metab_Occipital.dat
- sub-*_acq*-09_svs_slaser_cows7_mm_Occipital.dat
- sub-*_acq*-10_svs_slaser_cows12_metab_Occipital.dat
- sub-*_acq*-11_svs_slaser_vapor7_metab_Parietal.dat
- sub-*_acq*-12_svs_slaser_vapor7_mm_Parietal.dat
- sub-*_acq*-13_svs_slaser_cows7_metab_Parietal.dat
- sub-*_acq*-14_svs_slaser_cows7_mm_Parietal.dat
- sub-*_acq*-15_svs_slaser_cows12_metab_Parietal.dat
MAT FILE SPECTRA
Prefrontal Cortex
- Metabolite Spectrum - using COWS(7;236) Water Suppression
- Metabolite Water Scan - after COWS(7;236)
- Macromolecule Spectrum - using COWS(7;236) Water Suppression
- Macromolecule Water Scan - after COWS(7;236)
- Metabolite Spectrum - using COWS(12;626) Water Suppression
- Metabolite Water Scan - after COWS(12;626) Metabolite Scan
- Metabolite Spectrum - using VAPOR(7;626) Water Suppression
- Metabolite Water Scan - after VAPOR(7;626) Metabolite Scan
- Macromolecule Spectrum - using VAPOR(7;626) Water Suppression
- Macromolecule Water Scan - after VAPOR(7;626) Metabolite Scan
Posterior Frontal Lobe
- Metabolite Spectrum - using COWS(7;236) water suppression
- Metabolite Water Scan - after COWS(7;236)
- Macromolecule Spectrum - using COWS(7;236) water suppression
- Macromolecule Water Scan - after COWS(7;236)
- Metabolite Spectrum - using COWS(12;626) water suppression
- Metabolite Water Scan - after COWS(12;626) metabolite scan
- Metabolite Spectrum - using VAPOR(7;626) water suppression
- Metabolite Water Scan - after VAPOR(7;626) metabolite scan
- Macromolecule Spectrum - using VAPOR(7;626) water suppression
- Macromolecule Water Scan - after VAPOR(7;626) metabolite scan
Occipital Lobe
- Metabolite Spectrum - using COWS(7;236) water suppression
- Metabolite Water Scan - after COWS(7;236)
- Macromolecule Spectrum - using COWS(7;236) water suppression
- Macromolecule Water Scan - after COWS(7;236)
- Metabolite Spectrum - using COWS(12;626) water suppression
- Metabolite Water Scan - after COWS(12;626) metabolite scan
- Metabolite Spectrum - using VAPOR(7;626) water suppression
- Metabolite Water Scan - after VAPOR(7;626) metabolite scan
- Macromolecule Spectrum - using VAPOR(7;626) water suppression
- Macromolecule Water Scan - after VAPOR(7;626) metabolite scan
MRI FILES Each T1w MPRAGE brain scan was brain extracted using HDBET (https://github.com/MIC-DKFZ/HD-BET).
Before MRS: This is the T1w MPRAGE scan for voxel placement
- sub-*/anat/sub-*_T1w.nii.gz
After MRS: Second T1w MPRAGE scan performed within the same session for subject movement calculations after the MRS scans.
- derivatives/mrs_mat/sub-*_mat/sub-*_T1w.nii.gz
SOFTWARE and FILE TYPES Included
- COWS (Constrained optimized water suppression)
- SOFTWARE: https://inventions.techventures.columbia.edu/technologies/cows-water--CU21111
- Manuscript: Constrained optimized water suppression for 1H MR spectroscopy https://doi.org/10.1002/mrm.30550
- INSPECTOR Compatible -> .MAT
- SOFTWARE: https://inventions.techventures.columbia.edu/technologies/inspector-magnetic--CU17130
- Manuscript: INSPECTOR: free software for magnetic resonance spectroscopy data inspection, processing, simulation and analysis https://doi.org/10.1038/s41598-021-81193-9
- TWIX File format -> .DAT